Remote Control - Commands
NRP2
227User Manual 1173.9140.02 ─ 07
Note:
Every measured value displayed in the R&S NRP2 (also every point of a trace) is aver-
aged in the sensor via a series of samples. If a RANDom feed is selected, a random
value is taken from these samples and forwarded to the base unit as a measured
value. PEAK is the maximum of all samples in the measurement interval.
Suffix:
<[1]|2]>
.
Parameters:
<feed> string
see table 6-9 <string> values.
*RST: "POWer:AVERage"
CALCulate<[1]..8>:LIMit:BEEP <beep>
Switches the acoustic warning for limit violations on or off.
Parameters:
<beep> ON | OFF
*RST: OFF
Usage: Setting only
CALCulate<[1]..8>:LIMit:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
Usage: Event
CALCulate<[1]..8>:LIMit:CLEar:AUTO <auto>
Automatically resets the limit monitoring state and the internal counter for limit viola-
tions if any of the following events occurs:
INITiate[:IMMediate] is executed
INITiate:
CONTinuous ON is executed
MEASure? or MEASure:...? query is executed.
READ? or READ:...? query is executed.
CALCulate (Configuration of Calculate Blocks)