Remote Control - Commands
NRP2
228User Manual 1173.9140.02 ─ 07
Parameters:
<auto> ON | OFF | ONCE
ON
automatic reset enabled
OFF
automatic reset disabled
ONCE
automatic reset is enabled only once for the next occurrence of
the above events
*
RST: OFF
Usage: Event
CALCulate<[1]..8>:LIMit:FAIL?
Queries whether upper or lower limits have been exceeded. The status is reset (0 = no
limit violation) if one of the following events occurs:
The device is switched on
A reset is performed (*RST)
The
CALCulate<[1]..8>:LIMit:CLEar command is sent.
Usage: Query only
CALCulate<[1]..8>:LIMit:FCOunt?
Queries the number of limit violations that occurred. The counter is zeroed if one of the
following events occurs:
The device is switched on
A reset is performed (*RST)
The CALCulat<[1]..8>:LIMit:
CLEar command is sent.
Usage: Query only
CALCulate<[1]..8>:LIMit:LOWer[:DATA] <lower_limit>
Sets a lower limit for the measured values. The R&S NRP2 can respond when the val-
ues drop below this limit (see other CALCulate<[1]..8>:LIMit commands).
Parameters:
<lower_limit>
float_value
The value range depends on the current output unit of the calcu-
late block's measured value, see table below.
The parameter unit is determined according to table 6-11 Output
unit of measured value of calculate block.
*RST: 0 DBM or 0 DB
Default unit: Default unit: see table referred to above.
CALCulate (Configuration of Calculate Blocks)