Remote Control - Commands
NRP2
230User Manual 1173.9140.02 ─ 07
CALCulate<[1]..8>:LIMit:UPPer[:DATA] <upper_limit>
Sets an upper limit for the measured values. The R&S NRP2 can respond when this
limit is exceeded (see other CALCulate<[1]..8>:LIMit
commands).
Parameters:
<upper_limit> float_value
The value range depends on the current output unit of the calcu-
late block's measured value, as shown in table 6-10 Value
ranges, depending on the unit.
The parameter unit is determined according to table 6-11 Output
unit of measured value of calculate block.
*
RST: 0 DBM or 0 DB
Default unit: see table referred to above.
CALCulate<[1]..8>:LIMit:UPPer[:DATA]:POWer <power>
Sets a upper limit for the measured power. The R&S NRP2 can respond when this limit
is exceeded (see other CALCulate[1]..8>:
LIMit
commands).
Parameters:
<power> float_value; units: DBM | W | DBUV
Range: -120 DBM to 90 DBM
*RST: 0 DBM
Default unit:
DBM
CALCulate<[1]..8>:LIMit:UPPer[:DATA]:RATio <ratio>
Sets an upper limit for the measured power ratios. The R&S NRP2 can respond when
this limit is exceeded (see other CALCulate<[1]..8>:LIMit
commands).
Parameters:
<ratio> float_value, units: DB | DPCT | O
Range: -200 DB to 200 DB
*RST: 0 DB
Default unit:
DB
CALCulate<[1]..8>:LIMit:UPPer:STATe <state>
Switches the monitoring function for the upper limit on or off.
Parameters:
<state> ON | OFF
*
RST: OFF
Usage: Setting only
CALCulate (Configuration of Calculate Blocks)