Remote Control - Commands
NRP2
243User Manual 1173.9140.02 ─ 07
Parameters:
<display_mode> DIGital | ANALog | GRAPhical
DIGital
Measured values are displayed in numeric format. If the primary
channel is operated in Trace mode, the (scalar) measured val-
ues of the currently selected timegate are displayed here.
ANALog
Measured values are indicated by a marker on a scale. The val-
ues at the upper and lower scale end as well as the type of scal-
ing (linear or logarithmic) are set with the following :METer com-
mands.
GRAPhical
Measured values are plotted over time. This kind of presentation
is possible only if the primary channel is operated in Trace
mode.
*
RST: DIGital
DISPlay[:WINDow]<[1]..4>:METer:AUTO
<auto>
DISPlay[:WINDow]<[1]..4>:ANALog:AUTO <auto>
DISP:AUTO ONCE
automatically determines scaling for the analog display. The upper
and the lower limit value of the display is set as a function of the current measurement
data. Only the ONCE parameter is permissible; a query always supplies the value OFF.
Parameters:
<auto> ONCE | OFF
*RST: OFF
DISPlay[:WINDow]<[1]..4>:METer:LOWer <lower_limit>
DISPlay[:WINDow]<[1]..4>:ANALog:LOWer <lower_limit>
Sets the lower limit value of the analog scale.
The value range depends on the current output unit of the measured value of the win-
dow.
Table 6-12: Value range
Unit of measured value W DBM PCT DB
Value range for <float_value> 1e-18..1e18 -150..210 1e-18..1e22 -200..200
Unit: See table 6-10 Value ranges, depending on the unit.
Parameters:
<lower_limit> float_value
*RST: -90 DBM or -120 DB