Remote Control - Commands
NRP2
325User Manual 1173.9140.02 ─ 07
Parameters:
<key_stroke> <NR1> | <character_value> (see table)
<NR1>
System will be switched off SYSTem:KEY
17
<character_value>
System will be switched off SYSTem:KEY
POWER
Usage: Event
SYSTem:LANGuage <emulation>
Selects an emulation of a different command set. The R&S NRP2 is able to emulate
different power meters by changing the remote command set. As the list of supported
instruments is extended continuously, the emulations are documented in separate
application notes.
Parameters:
<emulation> character_data
NRP | NRP2 | HP436A | HP437B | HP438A | E4418B | E4419B
| N1911A | N1912A
Usage: Setting only
SYSTem:OPT <string>
Determines the user defined return value for the *OPT query. The maximum string
length is 128 characters.
SYSTem:OPT:AUTO <state>
Activates/deactivates return of the user-defined string for the *OPT query.
SYSTem:PRESet
Resets the R&S NRP2 to default (*RST) values. No difference is made between reset
and preset values (exceptions: INIT:CONT
, SENS:AVER:TCON, and
SENS:TRAC:AVER:TCON). For preset values see table 6-2.
Usage: Event
SYSTem:REBoot
Reboots the R&S NRP2.
Usage: Event
SYSTem