Instrument Functions
NRP2
62User Manual 1173.9140.02 ─ 07
all channels in the Cont Av mode (continuous average power)
absolute power measurement in dBm
one window per channel opened
offset: 0 dB
automatic filtering (normal mode)
(PRE)SET does not reset the following settings:
offset tables
address of the remote interface
SCPI error queue
selection of the beeper
brightness setting of the display
test limits for sensor test
SCPI command:
*RST*RST on page 198
(Default state for remote-control operation)
or
SYSTem:PRESet on page 325 (Default state for manual operation)
4.1.2 Frequency hardkey
FREQ is used to set the carrier frequency of the applied signal. This corrects any fre-
quency-dependent effects introduced by the sensor.
If the sensor is to reach the specified measurement accuracy, the signal frequency
must be entered.
4.1.2.1 Frequency dialog box
Fig. 4-1: Frequency dialog box
A | B | C | D
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix 1 for channel A, channel 2 for B, and so on.
Main Settings