Instrument Functions
NRP2
63User Manual 1173.9140.02 ─ 07
Frequency
Field for frequency entries with the units kHz, MHz or GHz.
Remote command:
[SENSe<[1]..4>]:FREQuency[:
CW] on page 269
4.1.3 Zero hardkey
ZERO starts the autozero function. Zeroing removes offset voltages from the analog
circuitry of the sensors, so that there are only low powers displayed when there is no
power applied.
Turn off all test signals before zeroing.
4.1.3.1 Zero dialog box
Fig. 4-2: Zero dialog box
The colored background of the text indicates that the function can also be started by
pressing the ZERO hardkey (again).
Zero All
Zeroes all connected sensors.
Remote command:
CALibration:ALL:ZERO:
AUTO on page 237
Zero A | B | C | D
Zeroes the selected channel.
Remote command:
CALibration<[1]..4>:ZERO:AUTO on page 237
4.1.3.2 Zeroing failure
If the power measured during zeroing is too large, the R&S NRP2 reports an error.
Main Settings
北京海洋兴业科技股份有限公司(证券代码:839145)