Instrument Functions
NRP2
82User Manual 1173.9140.02 ─ 07
Auto Config...
Opens the "Auto Filter" dialog box.
Fig. 4-17: Auto filter dialog box
A | B | C | D ← Auto Config...
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix 1 for channel A, channel 2 for B, and so on.
Mode ← Auto Config...
Selects the auto filter mode.
"Normal"
Selects the "Normal" autofilter mode.
"Fixed Noise"
Selects the "Fixed Noise" autofilter mode.
Remote command:
[SENSe<[1]..4>]:AVERage:COUNt:
AUTO:TYPE on page 265
Ref Timeslot ← Auto Config...
Specifies the timeslot from which the autofilter setting, i.e. the averaging number, is
derived for measurements in the "T'slot Av" mode. This parameter is not available in
the other measurement modes.
Remote command:
[SENSe<[1]..4>]:
AVERage:COUNt:AUTO:SLOT on page 265
Noise Content ← Auto Config...
Defines the proportion of intrinsic noise in the measured result (in dB). Specifically, this
value gives the permitted relative deviation from the mean which may not be exceeded
for 95% of the observation time.
Remote command:
[SENSe<[1]..4>]:
AVERage:COUNt:
AUTO:NSRatio on page 265
Max Settling Time ← Auto Config...
Specifies an upper limit for the settling time in the
"Fixed Noise"
mode.
Remote command:
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:MTIMe on page 264