Instrument Functions
NRP2
85User Manual 1173.9140.02 ─ 07
4.2.5.1 Sensor Γ correction dialog box
►
Open the Γ correction dialog box in the "Sensor "menu
If "Γµ correction" is activated, the measurement accuracy can be increased by specify-
ing magnitude and phase of source Γ.
Fig. 4-19:
Γ
Correction dialog box
A | B | C | D
Tabs for selecting the sensor.
Note: In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:
..., i.e. choose suffix 1
for channel A, channel 2 for B, and so on.
Source Γ Correction
Activates gamma correction in order to achieve higher measurement accuracy.
Remote command:
[SENSe<[1]..4>]:SGAMma:CORRection:
STATe on page 280
Source Γ Magnitude
Magnitude of the source reflection coefficient Γ.
Remote command:
[SENSe<[1]..4>]:SGAMma[:MAGNitude] on page 279
Source Γ Phase
Phase angle of the source reflection coefficient Γ in degrees.
Remote command:
[SENSe<[1]..4>]:
SGAMma:PHASe on page 279
4.2.6 Trigger settings
The trigger system is required for the "Trace & Statistics", "Timeslot Av", "Timegate
Av", and "Burst Av" measurement modes.
Data acquisition and parameters