Command Reference
R&S
®
ZNB/ZNBT
1308User Manual 1173.9163.02 ─ 53
Parameters:
<Type> POSitive | NEGative
*RST: NEGative
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:POSition <Type>
Specifies whether the trigger output pulse is sent before or after the selected sweep
interval (see TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval).
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<Type> BEFore | AFTer
BEFore
is appropriate if the device needs to be triggered before the data
is acquired (typical example: a power meter which is used for
power calibration)
AFTer
is appropriate if the device needs to be triggered just after data
acquisition (typical example: an external generator which must
be reconfigured to get ready for the next measurement)
*RST: AFTer
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
on page 1306
8.4 R&S ZVR/ZVABT Compatible Commands
The commands in this chapter are supported for compatibility with analyzers of the
R&S ZVR and R&S ZVAB family; they do not introduce any new functionality. For new
programs, it is recommended to use the commands in Chapter 8.3, "SCPI Command
Reference", on page 751.
CALCulate<Chn>:LIMit:CONTrol:DOMain.......................................................................1309
CALCulate<Chn>:LIMit:RDOMain:COMPlex................................................................... 1310
CALCulate<Chn>:LIMit:RDOMain:FORMat.....................................................................1311
CALCulate<Chn>:LIMit:RDOMain:SPACing.................................................................... 1311
CALCulate<Chn>:LIMit:LOWer:STATe............................................................................ 1311
CALCulate<Chn>:LIMit:UPPer:STATe.............................................................................1311
CALCulate<Ch>:PARameter:DEFine..............................................................................1312
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth........................................................ 1313
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DELTa:STATe................................................ 1314
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARGet........................................................ 1314
R&S ZVR/ZVABT Compatible Commands