The trigger duration must be shorter than the selected sweep interval (see TRIGger:
CHANnel<Ch>:AUXiliary<n>:INTerval).
If the trigger duration is too long, you can modify one of the following settings:
Reduce the trigger duration, if possible.
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<TrigOutDuration> Pulse width between 1 μs (1E-6 s) and 1 s.
*RST: 1 us
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
on page 1306
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval <Type>
Specifies whether the trigger output pulses are sent once every measurement point or
once every sweep.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<Type> POINt | SWEep
ACQuisition
A trigger out signal is generated for every data acquisition phase
*RST: SWEep
Example: See TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:OPOLarity <Type>
Selects the polarity of the output trigger pulses; see Figure 8-1.
Suffix:
<Ch>
.
Channel number
<n> Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.