Reference Manual
00809-0100-4809, Rev DA
Appendix A: Specifications and Reference Data
September 2015
131
Specifications and Reference Data
Configuration buttons
D4
(16)
Analog zero and span
DZ
(26)
Analog zero and span
Typical model number: 3051CFC D C S 060 N 065 0 3 2 A A 1 WC E5 M5
1. Select Configuration Buttons (option code D4 or DZ) or Local Operator Interface (option code M4) if local configuration buttons are required.
2. Available with primary element technology P only.
3. 10-in. (250 mm) and 12-in. (300 mm) line sizes not available with Primary Element Technology A.
4. For the 10-in. (250 mm) and 12-in. (300 mm) line size, the alignment ring must be ordered (Installation Accessories).
5. For 2-in. (50 mm) line sizes the Primary Element Type is 0.6 for Primary Element Technology Code C.
6. Available with Primary Element Technology A only.
7. Option HR5 configures the HART output to HART Revision 5. Option HR7 configures the HART output to HART Revision 7. The device can be field configured
to HART Revision 5 or 7 if desired. HART Revision 5 is the default HART output.
8. For local addressing and configuration, M4 (Local Operator Interface) is required.
9. Requires wireless options and engineered polymer housing. Available approvals are FM Intrinsically Safe, (option code I5), CSA Intrinsically Safe (option
code I6), ATEX Intrinsic Safety (option code I1), and IECEx Intrinsic Safety (option code I7).
10. Only available with C6, E2, E5, I5, K5, KB and E8 approval. Not available with GE, GM, SBS, DA0, M4, D4, DZ, QT, HR5, HR7, CR, CS, CT.
11. Only available with Wireless Output (output code X).
12. Not available with Product certifications options E8, K8, E5, K5, C6, K6, E7, K7, E2, K2, E3, KB, KD.
13. Available with Primary Element Technology C only.
14. For Annubar option A, consult factory for pipe schedules other than schedule 40.
15. Available with Primary Element Technology C or P only.
16. Only available with HART 4-20 mA Output (output code A).
17. Materials of Construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental
limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
18. Dust approval not applicable to output code X.
19. Nonincendive certification not provided with Wireless output option code (X).
20. Not available with Wireless output (output code X).
21. Only available with product certifications E7, E8, I1, I7, IA, K7, K8, KD, N1, N7.
22. Not available with output code F - F
OUNDATION Fieldbus or Wireless output (output code X) or Low Power (output code M).
23. The T1 option is not needed with FISCO Product Certifications, transient protection is included with the FISCO Product Certification code IA, IB, and IE.
24. Only valid with F
OUNDATION Fieldbus (output code F).
25. The V5 option is not needed with the T1 option; external ground screw assembly is included with the T1 option.
26. Only available with 4-20 mA HART Output (output code A) and Wireless output (output code X).
Table 7. Rosemount 3051CFC Compact Flowmeter Ordering Information
The Standard offering represents the most common options. The starred options () should be selected for best delivery.
__The Expanded offering is subject to additional delivery lead time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rosemount 3051SFC_A and is the answer not in the manual?

Rosemount 3051SFC_A Specifications

General IconGeneral
BrandRosemount
Model3051SFC_A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals