MPM-210H OPERATION MANUAL
36 / 59
measuring mode for each channel.
Refer to the 5.5.7 for the relation between the gain setting
and the input power.
This command is not available for the MPM-215.
Syntax
DLEV<wsp><value1>,<value2>,<value3>
Parameters
<value1>: Module 0, 1, 2, 3, 4, 5
When setting the <value 1> to 5, all channels and
module is set to the same wavelength.
<value2>: Channel 1, 2, 3, 4
<value3>: Gain 1, 2, 3, 4, 5
In case of the MPM-213 (current meter), the available
range is 1 to 4.
Default
value3: 5
Example
DLEV 0,1,5
16. DLEV?
Description
Check TIA Gain of CONST1, SWEEP1, FREERUN, AUTO1
measuring mode for each channel.
Refer to the 5.5.7 for the relation between the gain setting
and the input power.
When using the READ? command, the manual mode
(AUTO1) of the range mode is only available.
This command is not available for the MPM-215.
Syntax
DLEV? <wsp><value1>,<value2>
Parameters
<value1>: Module 0, 1, 2, 3, 4
<value2>: Channel 1, 2, 3, 4
Response
<value>
value: 1, 2, 3, 4, 5
Example
DLEV? 0,1
Response: 5
17. AVG
Description
Set up Average Time.
Syntax
AVG<wsp><value>
Parameters
<value>: 0.01 ~ 10000.00