Unit: ms
Resolution: 0.01(10us)
Default
value: 5
Example
AVG 5
18. AVG?
Description
Check Average time.
Syntax
AVG?
Response
<value>
value: 0.01 ~ 10000.00
Unit: ms
resolution: 0.01(10us)
Example
AVG?
Response: 5
19. FGSAVG
Description
Set Average Time. (This command is same operation as
“AVG”.)
Syntax
FGSAVG<wsp><value>
Parameters
<value>: 0.01 ~ 10000.00
Unit: ms
Resolution: 0.01(10us)
Default
value: 0.1
Example
FGSAVG 0.1
20. FGSAVG?
Description
Check Average time. (This command is same operation as
“AVG?”.)
Syntax
FGSAVG?
Response
<value>
value: 0.01 ~ 10000.00
Unit: ms
resolution: 0.01(10us)
Example
FGSAVG?
Response: 0.1