MULTI-PORT OPTICAL POWER METER MPM-210H
39 / 59
Example
AUTO?
Response: 1
25. DAUTO
Description
Set the power range (Auto or Manual) for each channels.
It is available for the “READ?” command.
Syntax
DAUTO<wsp><value1>,<value2>
Parameters
<value1>: Module 0,1, 2, 3, 4, 5
When setting the <value 1> to 5, all channels and
module is set to the same wavelength.
<value2>: Range mode
0 Manual range
1 Auto range
Default
value2: 1
Example
DAUTO 0,1
26. DAUTO?
Description
Check the power range (Auto or Manual) for each
channels.
It is available for the “READ?” command.
Syntax
DAUTO?<wsp><value>
Response
<value>: Range mode
0 Manual range
1 Auto range
Example
DAUTO? 0
Response : 1
27. READ?
Description
Execute optical power or electrical current measurement
and check the result. If sending the “READ” command
during logging mode, logging mode stops.
Syntax
READ?<wsp><module>
Parameter
<module>
Module Number: 0,1,2,3,4
Response
<value1>,<value2>, <value3>,<value4>