118 Operating Instructions Combics CAIXS2
Konfigurationsbeispiel Combics 2 EX - FM
100- 240V
Sicherer Bereich
(NON Ex)
YPS02-XUR
inkl.
(3m)
inkl.
(3m)
YDI05-Z
mit A25
z.B. Drucker
YDP14-OCE
COM 1
RS422
M54
(x m)
inkl.
(0,3m)
M57
(1,5 m)
RS232
FM Class I, Div 1
Anschlüsse
Accessories
Power Supply
Item
Order No.
AC adapter, for use in explosive atmospheres
100–240 V 14-pin round plug (30 cm) ATEX YPSC01-X
USA YPS02-XUR
Canada YPS02-XKR
AC adapter outside explosive atmospheres ATEX YPSC01-Z
100-240 V range USA/Canada YPS02-ZKR
AC adapter outside explosive atmospheres 24 V ATEX YPS02-XV24
External battery for installation in explosive area YRB02-X
USA/Canada
Configuration example
Ex area Safe area
(Non Ex)
Installation Instructions | Betriebsanleitung
Sartorius YPSC01-X
Configurable power supply and communication system
for use inside hazardous areas
Konfigurierbares Netzgerät und Kommunikationssystem
für den Einsatz innerhalb des explosionsgefährdeten Bereichs
Sartorius YPSC01-Z
Configurable power supply and communication system
for use outside hazardous areas, only
Konfigurierbares Netzgerät und Kommunikationssystem
nur für den Einsatz außerhalb des explosionsgefährdeten Bereichs
98647-004-37
Installation
Connecting the Device to AC Power
!Before startup, make sure that the power cable is properly connected to the power
outlet. In particular, the protective conductor must be connected to the housing
of the power supply.
Connect all devices via their equipotential bonding conductor terminals to the
equipotential bonding conductor. Installation must be carried out properly by
trained personnel and according to the standards of the technology.
!The system should only be operated for the first time when it is certain that the
area is not potentially explosive. If deviations are evident during startup due to
transport damage, the system should be disconnected from the power supply and
Service should be contacted (no display, no backlighting despite weight value
display, no reproducibility of the weight value, no stability in the display, etc.)
!Cabling:
– Both cables should be laid fixed
– Flexible cabling with screws available upon request
t Check the supply voltage/frequency and plug design.
Contact your Sartorius dealer if they do not match.
!When operating a balance in an explosive area:
– Follow all current standards and regulations for the installation of devices in the
explosive area.
– Device installation should be carried out by trained personnel and checked
properly.
!Before startup, connect the linked devices to a central equipotential busbar.
h
Only use cables and cable lengths approved by Sartorius,
cables not provided by Sartorius are the responsibility of the operator.
6 YPS02-X.. Installation Instructions
Installation
Page 20
Power Supplies for Signum® 3 Ex Supreme
ATEX Zone 20/1/21/2/22 Safe Area: YPSC01-Z
YP
SC01-X Associated Intrinsically Safe (AIS)
to connect to
ATEX Zone 20/1/21/2/22
FM Class I,II,III Div 1,2 Safe Area: YP
S02-ZKR
YPS02-XUR AIS to connect to
C
SA Class I,II,III Div 1,2 FM /CSA Class I,II,III Div 1,2
YP
S02-XKR
ATEX Zone 20/1/21/2/22 Safe Area: YPS02-XV24
YP
S02-XV24 AIS to connect to
ATEX Zone 20/1/21/2/22
ATEX Zone 20/1/21/2/22
FM Class I,II,III Div 1,2
YR
B02-X
Power Supplies
100 - 240 V
Rechargeable
B
attery
Power Supply
24V
26. November 2012
98647-003-43
Operating Instructions | Betriebsanleitung
Sartorius
YRB02-X
Ex Rechargeable Battery Pack (External Explosion-protected Rechargeable Battery Pack)
Ex-Akku-Pack (externer wiederaufladbarer Ex-geschützter Akku-Pack)
YPSC01-X