EasyManua.ls Logo

Sartorius Combics 2 CAIXS2 - Counting Z; Average Piece Weight

Sartorius Combics 2 CAIXS2
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Combics CAIXS2 51
Operation
Counting Z
With the Counting application, you can determine the number of parts to each
have approximately equal weight (menu APPLIC.1).
Features Save the reference weight “wRef" from the weighing platform.
Enter the reference sample weight “wRef" using the keypad.
Enter the reference sample quantity “nRef" using the keypad.
Automatic average piece weight updating.
Counting with two weighing platforms.
Activate Info mode with the I key.
Toggle the display between quantity and weight using the W key.
»Define the level of accuracy (display resolution) applied when a calculated
reference sample quantity is saved.
Automatic taring of container weight.
Setting: Applic./aut.tare, menu item 3.7.
Automatic initialization when the scale is switched on. The indicator is
initialized with the most recently used values for reference sample quantity
“nRef” and reference sample weight “wRef.” Setting: Applic./aut.strt,
menu item 3.8.
Exit Application, Delete Parameters
The value of the reference sample weight in the reference memory remains active
until it is deleted using the c key, or is overwritten, or until the application is
changed. The reference sample weight also remains saved after the scale is turned
off.
You can assign different functions to the c key to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application. Setting: Applic. / CLER.CF menu
item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value. Setting:
Applic./TARE.FNC. Menu item 3.25.1 (factory settings). A tare value entered
manually overwrites any stored tare value (weight value). If you enter a tare value
manually, a tare value (weight value) stored later overwrites the manually entered
value. Setting: Applic./TARE.FNC menu item 3.25.2.
Restore factory default settings: Applic./Reset menu item 9.1.
Average Piece Weight
Before the quantity on the platform can be calculated, the average piece weight
must be entered in the application. There are several ways to enter this value in the
program:
Calculating the Reference Piece Weight
Place the number of parts defined as the reference sample quantity on the
weighing platform and calculate the average piece weight by pressing the
O key.
or
Place any number of parts on the connected weighing platform, enter the
quantity using the keypad and select and calculate by pressing the r key.
How the reference weight is calculated depends on the application setting for
resolution. The value is either rounded off in accordance with the display
resolution, or saved with 10-fold or 100-fold resolution.

Table of Contents

Related product manuals