22 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Configuring Weighing Platforms
Service mode
Purpose The Service mode enables access to additional menu items in the Setup menu
(setup) which are not displayed when the Service mode is not active. The most
important calibration and adjustment work for the indicator and for the connected
weighing platform can be carried out in the Service menu, e.g. ADC configuration.
When the Service mode is active, an “S" is shown in the top right-hand corner of
the display. To deactivate the Service mode, restart the indicator (turn the indicator
off and back on again).
In Service mode, the Setup menu is expanded with the following parameters after
entering the user password:
S-DATE for entering the next service date
SER.No. for entering the device serial number
MODEL with the model description
S-SQMIN
ALIB.MEM for deleting the alibi memory.
The Setup menu for WP1 and WP2can be extended to include the following setting
options to confugure the weighing platforms:
Param1
cal.adj calibration, adjustment 1.9
lIN.INT internal linearization 1.9.5
lIN.ext. external linearization with default weights 1.9.6
lIN.e.usr external linearization with user-defined weights (entered under 1.18) 1.9.7
SET.PREL. set the preload (not for use in legal metrology) 1.9.8
DEL.PREL delete the preload (not for use in legal metrology) 1.9.9
HND.XT.G/CAL.ADJ enter the adjustment and linearization weights 1.18
CAL.ADJ.1 enter adjustment weight 1.18.1
LIN. WT.1 enter lin. weight 1 1.18.2
LIN. WT.2 enter lin. weight 2 1.18.3
LIN. WT.3 enter lin. weight 3 1.18.4
LIN. WT.4 enter lin. weight 4 1.18.5
ADJ.W/O.W adjust without weights (enter the characteristic data of the load cells) 1.19
NOM.LOAD. nominal load 1.19.1
RESOLUT resolution (only visible for older ADCs)* 1.19.2
SENSIT.1 sensitivity in mV/V for cell 1 (or average value for all cells) 1.19.3
SENSIT.2 sensitivity in mV/V for cell 2 1.19.4
SENSIT.3 sensitivity in mV/V for cell 3 1.19.5
SENSIT.4 sensitivity in mV/V for cell 4 1.19.6
ZER.POIN zero point or offset of system in mV/V 1.19.7
SAVE. save values for 1.19 1.19.8
GEOG.DAT adjustment location (geograph. data; or alternatively the gravitational acceleration
at the place of installation) 1.20
LATITUD latitude in degrees 1 20.1
ALTITUD elevation in meters above sea level 1 20.2
GRAVITY. gravitational acceleration 1 20.3
SAVE. save values for 1.20 1 20.4
ADC settings (see Analog/Digital Converter) 11
Apply the serial number of the IS weighing platform 12.1
Apply the serial number 12.1.1
Inactive (standard WP) 12.1.2