Operating Instructions Combics CAIXS2 69
Operation
Checkweighing H
With this application, you can check whether the sample on the weighing platform
matches a target value or lies within a given tolerance range. Checkweighing also
makes it easy to fill sample materials to a specified target weight.
Selection and settings in the Applic. / Applic.2/ CHECK.WG menu.
Features – Enter the nominal or target weight (set point) and the tolerance range
delimiters either using the keypad or by saving the weight value of a load on
the platform.
– Enter the tolerance limits as absolute values (Min and Max), as a percentage
deviation from the target or as a relative deviation from the target.
Setting: Applic./Applic.2/CHECK.WG/CHECK.RG menu item 4.5.
– The target value can be taken over as a weighed value from a weighing
platform, and the upper and lower tolerance limits are defined as a percentage
deviation from the target value (setting 4.5.2). The following percentages can
be selected as the deviation: 0.1%, 0.2%, 0.5%, 1%, 1.5%, 2%, 3%, 5% or 10%,
selection using the r key.
– The target value, lower tolerance limit (minimum) and upper tolerance limit
(maximum) can applied as weighed values from the weighing platform
(menu item 4.5.1).
– The target value can be applied as weighed values and via asymmetrical percent
limits (menu item 4.5.3).
– The target value can be applied as weighed values and via relative weight limits
(menu item 4.5.4).
– Target value and tolerance limits checked during input; values must conform
to: upper limit > target > lower limit > 1 digit.
– Checkweighing range: either 30% to 170% of the target, or from 10% to
infinity.
– Results are shown on the main display, as a bar graph and LED as well as sent
to control output ports for further processing.
– Toggle the main display between weight and tolerances limits by pressing the
w key. If the weight in the readout is outside the tolerance range, “LL" (too
low) or “HH" (too high) is displayed.
– Activate Info mode with the I key.
– Automatic results printout (Applic./Applic.2/CHECK.WG/CHECK.RG menu
item 4.6).
– Automatic taring of container weight (APPLIC./ aut.tare menu item 3.7).
– Automatic initialization when you switch on the scale with most recently saved
application data (APPLIC./aut.Strt menu item 3.8).
You can assign different functions to the c key to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting: APPLIC/CLER.CF menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: APPLIC./TARE.FNC menu item 3.25.1 (factory setting).