EasyManua.ls Logo

Sartorius Combics 2 CAIXS2 - Target value; Preparation

Sartorius Combics 2 CAIXS2
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70 Operating Instructions Combics CAIXS2
Operation
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic./TARE.FNC menu item 3.25.2.
Restore factory default settings: Applic./Reset menu item 9.1.
Target value Checkweighing entails comparing the current weight value to a defined target
value. You can enter the value for this target using the keypad, or by saving the
weight value indicated. You can also define upper and lower tolerance limits based
on this target. You can do this by:
entering absolute values using the keypad or placing the desired amount of
weight on the platform and saving the value
or
by entering each value using the keypad as a percentage deviation of the target
weight
or
by entering each value as an asymmetrical percentage deviation of the target
weight that is selected via the keypad or using the r key
or
by entering a relative weight deviation from the target weight via the keypad.
The value remains valid until deleted by pressing the c key or until overwritten
by a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC./APPLIC.2/CHECK.WG menu.
Available Parameter Settings * = Factory setting
MIN.INIT Minimum load for initialization 3.5
1 digit 1 scale interval 3.5.1*
2 digit 2 scale intervals 3.5.2
5 digit 5 scale intervals 3.5.3
10 dig. 10 scale intervals 3.5.4
20 dig. 20 scale intervals 3.5.5
50 dig. 50 scale intervals 3.5.6
100 dig. 100 scale intervals 3.5.7
200 dig. 200 scale intervals 3.5.8
500 dig. 500 scale intervals 3.5.9
1000 d 1000 scale intervals 3.5.10
Aut.strt Automat. Automatic start of applications when you switch on
the device with most recently saved application data 3.8
AUTOMAT Automatic (on) 3.8.1
MANUAL Manual (off) 3.8.2
TARE.FNC Tare function 3.25
Normal Can add a preset tare if tare value
is available; however no tare
function possible 3.25.1*
Special When a preset tare is entered, the tare
value is deleted; however, tare function
activation is possible 3.25.2
CHECK.RG Checkweighing range 4.2
30- 170% 30 to 170% 4.2.1*
10-MAX.L 10% to infinity 4.2.2

Table of Contents