36 Operating Instructions Combics CAIXS2
Operation
Operation
Weighing
This application is always available during operation.
Features: Zeroing by pressing (
Storing the weight on the platform as a tare by pressing )
Taring container weight automatically
Using a 10-key keypad to enter tare weight
Deleting tare values by entering 0 and ) / c and )
Toggling the display using the Fn key between:
– 1st and 2nd weight unit
– SQmin
Configuring the k key function in the “Fn key" Setup menu
10-fold increased resolution using the K key
Toggling between the gross or net value using L Weighing with two
weighing platforms
Individual numeric ID codes for weight values
Printing weight value:
GMP printout
Automatic printout
Automatic data output (see Data Interfaces chapter)
Automatic Taring (APPLIC menu item 3.7):
When the menu item is active (3.7.2), the first weight on the scale that exceeds the
preset minimum load is stored in the tare memory at stability.
The scale returns to the initial state when the load on the scale is less than 50% of
the minimum load.
Minimum load for automatic taring and automatic printing
(menu item 3.5):
You can set the following for the minimum load:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits" here refer to the scale intervals for the connected weighing platform.
If the interval is 1 g and 1000 digits are required, the minimum load is 1000 g
(1000 intervals).
If the weighing platform interval is 5 g and the same number of digits as above is
required, the minimum load is 5000 g.
When the load exceeds the minimum load limit, the weighing platform is tared
automatically and/or a report printout is generated automatically; however, this
requires the corresponding menu items to be active for automatic taring (menu
item 3.7.2) and for automatic printing (menu 7.15.2).