56 Operating Instructions Combics CAIXS2
Operation
Neutral Measurement Z nM
With this application you can measure the length, surface and volume of parts that
have roughly the same specific weight. The o symbol is displayed as the unit (menu
APPLIC.1).
Features Save the reference weight “wRef" from the weighing platform.
Enter the reference weight “wRef" using the keypad.
Enter the factor for calculation “nRef” using the keypad.
Measuring with two weighing platforms.
Activate Info mode with the I key.
Toggle the display between measurement and weight using the w key.
The level of accuracy (display resolution) can be set when the calculated
reference weight is applied.
Automatic taring of container weight.
Setting: Applic./aut.tare, menu item 3.7.
Automatic initialization when the scale is switched on. The indicator is
initialized with the most recently used values for reference sample quantity
“nRef” and reference sample weight “wRef.”
Setting: Applic./aut.strt, menu item 3.8 .
Exit Application, Delete Parameters
The value of the reference sample weight in the reference memory remains active
until it is deleted using the c key, or is overwritten, or until the application is
changed.
The reference sample weight also remains saved after the scale is turned off.
You can assign different functions to the c to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting: APPLIC/CLER.CF menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: APPLIC./TARE.FNC menu item 3.25.1 (factory setting).
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic./TARE.FNC menu item 3.25.2.
Restore factory default settings: Applic./Reset menu item 9.1.
Reference Weight
In order to calculate the length, surface or volume of a given sample, the average
weight of a reference quantity of the sample must be known (in the example below,
the reference is 1 meter of electrical cable). There are different ways to enter the
reference weight:
Calculating the Reference Weight
- Place the reference quantity (defined by the calculation factor) on the
connected weighing platform and calculate the reference sample weight by
pressing the O key.
or
- Place any amount of the sample material on the connected weighing platform,
enter the calculation factor through the keypad, and press the r key to
calculate the reference sample weight.