Operating Instructions Combics CAIXS2 33
Configuring Weighing Platforms
Set preload
Setup information Setting the preload when weighing in legal metrology is only possible using the
“Zero at Power On" menu item.
The “Set Preload" function (menu item 1.9.8) must be allocated to the J key.
3
Once the preload has been set, the J key must be reallocated back to its original
function in the Setup menu, e.g. external calibration/adjustment with default
weights (Setup menu item 1.9).
Procedure
( t Zero the weighing platform.
t Place the preload weight on the weighing platform.
J t Start the “Set Preload" function.
y After a short period of time the preload will be applied and the indicator will
automatically switch back to weighing mode.