Operating Instructions Combics CAIXS2 33
Configuring Weighing Platforms
Set preload
Setup information – Setting the preload when weighing in legal metrology is only possible using the
“Zero at Power On" menu item.
– The “Set Preload" function (menu item 1.9.8) must be allocated to the J key.
3
Once the preload has been set, the J key must be reallocated back to its original
function in the Setup menu, e.g. external calibration/adjustment with default
weights (Setup menu item 1.9).
Procedure
( t Zero the weighing platform.
t Place the preload weight on the weighing platform.
J t Start the “Set Preload" function.
y After a short period of time the preload will be applied and the indicator will
automatically switch back to weighing mode.