Loading...
140 Operating Instructions Combics CAIXS2
External Data Interface
External Connections and
Data Transfer
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
1
9
66015-741-50
CAIXS2
YPS02-X.R
EX Power Supply
(100 -240 Vac)
YPS02-XV24
EX Power Supply
(24 Vdc)
YRB02-X
EX Rechargeable
Battery Pack
YPS02-Z.R
EX Power Supply
(100 -240 Vac)
YPS02-XV24
EX Power Supply
(24 Vdc)
Power Supply
3m*
0.2m
2m*
20m*
5m*
Notes:
*: Customer-made product: up to 20m
**: Options up to 40m
5m*
5m
5m
3m
3m
YPSC01-X
EX Power Supply
(100 -240 Vac)
3m**
3m
YPSC01-Z
EX Power Supply
(100 -240 Vac)
20m**
3m
Non-hazardous areaHazardous area
DC Input Cable
8-contact
male
connector
External Connections and
Data Transfer
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
2
9
66015-741-50
CAIXS2
Option A16:
CAPXS.-... /
IUXS4-... /
IFXS4-...
weighing platform
CAPXS.-... /
IUXS4-... /
IFXS4-...
weighing platform
ADU
DMS Frontend Board
(Option A15; default)
WP Board
Alternative
YCC02-XR14M6
connecting cable
Weighing Point (WP) Connection
6m
Option A19:
YCC02-XR14M6
connecting cable
6m
Or
Option M20:
Option M26:
Non-hazardous areaHazardous area
Option X2:
Separable connection
3m or 6m (depends on model)
Alternately
connectable
IS......-.X...
weighing module
or
FC/FCA......-.X...
with
RS-232
data interface
IS......-.X...
weighing module
or
FC/FCA......-.X...
with
RS-485
data interface
J102
WP Board
J101
RS232
RS485

Table of Contents

Related product manuals