Operation
Configuring Printouts
Purpose You can individually define each measurement printout. This should be carried out
after setting the applications since some data in the printout is application-
dependent.
In the “Print parameters" menu, single, component, and total data records can be
configured, which contain the available print items for the respective applications.
Using the total data record for “Totalizing" and “Net-total formulation"
applications, you can define which parameters are printed using the c key.
Features Six lists each with a max. length of 30 print items
– Single printout Printer 1
– Component printout Printer 1
– Total data printout Printer 1
– Single printout Printer 2
– Component printout Printer 2
– Total data printout Printer 2.
Single, component, and total data records can be configured separately.
Print single printout: p key.
Auto printout of application when Setup menu is activated:
– Animal weighing (averaging)
– Checkweighing
– Classification.
Print component printout:
Totalizing/Net-total formulation with the O
key: Applic./Applic.3/TOTALIZ printout: Component printout.
Print total data printout:
For selected application Totalizing/Net-total formulation with c key.
When switching to another application in Setup, only the application-
dependent printout lists are deleted. The other printout lists remain saved.
Print items can be deleted individually: Press and hold the ( key.
Print items “Form Feed" for record footer:
Move to the next label start for printer type: YDP14IS: “Label" and YDP04IS,
setting “Label, manual form feed".
ISO/GMP-compliant printout: The Setup menu configuration under
“ISO/GMP-compliant printout” is also active for configured printouts.
Preparation t Open Menu mode (see page 35).
k k ... t Select the Setup menu.
k k ...) t Select and open the Printsubmenu.
k k ...) t Select and open the PROTOC.submenu.
Available parameter settings
PROTOC. Protocol 7
HEADLIN. Header and ID header input 7.4
QTY.1 Quantity interface 1 7.5
INDIV. 1 Standard interface 1 7.6
COMPON.1 Component interface 1 7.7
TOTAL 1 Result interface 1 7.8
QTY.2 Quantity interface 2 7.9
INDIV. 2 Standard interface 2 7.10
COMPON.2 Component interface 7.11

Table of Contents

Related product manuals