28 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Configuring the A/D Converter (ADC)
The weighing platform must already be connected.
Opening the Menu Access Switch
The menu access switch is located on the back of the indicator, behind the cover.
t Remove the cap.
t Slide the switch to the right (= “open" position).
Menu Access Switch/Calibration Switch Cover
e t Switch off and restart the device.
t While all segments are lit, press the ( key briefly.
y ADC-CON appears briefly on the display, followed briefly by S-CODE.
y The cursor flashes on the display.
t Enter the service password (see Appendix).
) t Confirm your entry using the ) key.
y The device is in Service mode. This can be recognized by the small S in the top
right of the display.
t Select the weighing platform to be configured, using the k key to switch to
WP-2 if required.
) t Confirm your selection using the ) key.
t Select the Configuration mode using the k key: STANDRD or VERIF.
t Carry out ADC configuration (see menu tree).
t Once you have completed the configuration, save the data using the SAVE
menu item.
y The indicator will restart automatically.