Operating Instructions Combics CAIXS2 27
Configuring Weighing Platforms
Minimum load (min. load) When “Standard configuration” is used, this menu item is not displayed.
The minimum load of the connected weighing platform is entered under this menu
item. The minimum load for scales of class l is 20 e and 10 e for class m.
Attention: The function of the minimum load setting is to warn operators that
below this limit, the summation of tolerances might lead to significant measure-
ment errors. In Germany, for example, initial weights below the minimum load are
not allowed.
Available weight units WT.UNIT Menu item 11.7
This menu item is used to select the weighing units that have been cleared for use
in weighing. All units marked with a circle (o) have been cleared for use, multiple
selection is possible.
Calibration/Adjustment unit Cal.Unit Menu item 11.8
This menu item is used to select the weighing unit that must be used for a
calibration/adjustment. The selected unit is then valid as a calibration/adjustment
unit even when a different unit is used during normal weighing operation.
Save parameters SAVE Menu item 11.10
The ADC configuration data is saved by selecting Menu item 11.10.1.