82 Operating Instructions Combics CAIXS2
Operation
Totalizing s
With this application, you can add weights to the totalizing memory. In addition
to weight values, the number of separate values added to memory is also saved
(APPLIC.3 menu).
Features – You can weigh up to 999 items.
– Save values automatically: Simultaneous saving of net values and calculated
values (if available).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
– Save weight values and calculated values from either Application 1 (for
example, Counting, Weighing in Percent) or Application 2 (Checkweighing).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.22.
– Current transaction number displayed in the number display (indicating the
transactions already added).
– Weighing in up to a defined target, with the totalization memory content
+ current weight displayed in the text lines.
– Save weight values manually or automatically.
– Accurate calculation of total of weight values from two weighing platforms.
– Activate Info mode with the I key .
– Automatic printout when value saved.
– Automatic taring of container weight.
Setting: Applic./aut.tare, menu item 3.7.
Exit Application, Delete Parameters
The value of the totalizing memory remains valid until deleted by pressing the
c key.
You can assign different functions to the c key to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting: APPLIC/CLER.CF menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: APPLIC./TARE.FNC menu item 3.25.1 (factory setting).
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic./TARE.FNC menu item 3.25.2 .
Restore factory default settings: Applic./Reset menu item 9.1.
The Signum has a totalizing memory for totalizing individual net and gross values.
You can save weight values in totalizing memory manually or automatically.
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
– Save value manually by pressing the O key.
The value taken from the active platform is added to the value already saved in
the totalization memory and the transaction counter value is increased by one.
When a value is added manually, the program does not check whether the
platform has been unloaded since the last time the O key was pressed.