EasyManua.ls Logo

Sartorius Combics 2 CAIXS2 - Totalizing s; Features

Sartorius Combics 2 CAIXS2
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
82 Operating Instructions Combics CAIXS2
Operation
Totalizing s
With this application, you can add weights to the totalizing memory. In addition
to weight values, the number of separate values added to memory is also saved
(APPLIC.3 menu).
Features You can weigh up to 999 items.
Save values automatically: Simultaneous saving of net values and calculated
values (if available).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
Save weight values and calculated values from either Application 1 (for
example, Counting, Weighing in Percent) or Application 2 (Checkweighing).
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.22.
Current transaction number displayed in the number display (indicating the
transactions already added).
Weighing in up to a defined target, with the totalization memory content
+ current weight displayed in the text lines.
Save weight values manually or automatically.
Accurate calculation of total of weight values from two weighing platforms.
Activate Info mode with the I key .
Automatic printout when value saved.
Automatic taring of container weight.
Setting: Applic./aut.tare, menu item 3.7.
Exit Application, Delete Parameters
The value of the totalizing memory remains valid until deleted by pressing the
c key.
You can assign different functions to the c key to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting: APPLIC/CLER.CF menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: APPLIC./TARE.FNC menu item 3.25.1 (factory setting).
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic./TARE.FNC menu item 3.25.2 .
Restore factory default settings: Applic./Reset menu item 9.1.
The Signum has a totalizing memory for totalizing individual net and gross values.
You can save weight values in totalizing memory manually or automatically.
Setting: Applic./APPLIC.3 /TOTALIZ menu item 3.16.
Save value manually by pressing the O key.
The value taken from the active platform is added to the value already saved in
the totalization memory and the transaction counter value is increased by one.
When a value is added manually, the program does not check whether the
platform has been unloaded since the last time the O key was pressed.

Table of Contents