Operating Instructions Combics CAIXS2 83
Operation
Value saved automatically when the weighing platform is stable and the
defined minimum load is exceeded.
If the defined minimum load is not exceeded, you can save the item manually
by pressing the O key. Regardless of these settings, the current value cannot be
saved automatically unless the platform is unloaded before the next sample is
placed on it. The weighing platform is considered to be unloaded when the
load is less than 50% of the minimum load.
The number of items added to memory is shown in the number display.
Press the c key to clear the totalizing memory. A printout is automatically
generated.
With two weighing platforms connected, you can add values from both platforms
to the totalizing memory. The displayed result is accurately calculated in the active
weight unit.
Example: When you add 1.243 g (determined on a weighing platform with three
decimal places) to 1.4 g (determined on a platform with 1 decimal place), the
display shows 2.643 g.
Preparation t Open the APPLIC./APPLIC.3/TOTALIZ. menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 scale interval 3.6.1*
2 digit 2 scale intervals 3.6.2
5 digit 5 scale intervals 3.6.3
10 dig. 10 scale intervals 3.6.4
20 dig. 20 scale intervals 3.6.5
50 dig. 50 scale intervals 3.6.6
100 dig. 100 scale intervals 3.6.7
200 dig. 200 scale intervals 3.6.8
500 dig. 500 scale intervals 3.6.9
1000 d 1000 scale intervals 3.6.10
AUTO.SAV Autosave 3.16.
OFF Off 3.16.1*
ON On 3.16.2
PRT.SAV. Individual/Component printout when saved 3.17.
OFF Automatic printing off 3.17.1
ON Print the entire standard print
configuration every time with the
O key 3.17.2*
VAL.FROM Source of data for autosave 3.22.
Appl. 1 Application 1 3.22.1*
Appl. 2 Application 2 3.22.2
SAV.VAL. Save value 3.23.
NET Net 3.23.1*
CALCUL. Calculation 3.23.2
NET+CAL Net and Calculated 3.23.3
Printout You can configure whether a printout is generated automatically when a weight
value is stored in the totalizing memory or manually by pressing the p key.
Setting: Applic./Applic.3/TOTALIZ menu item 3.17.
You can print manually by pressing the p key (single printout): 3.17.1.
Component log (single printout of an item): 3.17.2.
The total data record is printed when you clear the totalizing memory (by pressing
the c key).