8 Operating Instructions Combics CAIXS2
Getting Started
Getting Started
1.) Connect weighing platform to the indicator.
2.) CAIXS2 Indicators have an intrinsically safe data interface which can be
connected to a computer (or any other peripheral device) using a barrier
(e.g. YDI05-Z).
3.) Connect the AC adapter.
4.) Configure the analog/digital converter (ADC).
5.) Carry out an alignment: Adjustment.
Connecting Weighing Platforms
You can connect any intrinsically safe, analog Sartorius platform to your CAIXS2
Indicator. Refer also to the Verification of Intrinsic Safety, the EC Type Examination
Certificate for the CAIXS2 and the load cell or analog weighing platform to be
connected.
3
The load cell should be connected by a certified technician who has received
specialized training from Sartorius. Any installation work that does not conform
to the instructions in this manual results in forfeiture of all claims under the
manufacturer’s warranty.
3
Make sure the CAIXS2 is disconnected from the power before starting any
connection work.
t Set up the weighing platform (see operating instructions for the weighing
platform).
t Place the cable from the weighing platform next to the indicator.
t Open the Combics indicator:
Loosen the ten cap nuts on the front panel. Remove the front panel.
t Remove the cover plate by removing the 4 screws.
Connection terminals are now visible.
Installing Connection and Interface Cables
3
The cable gland (IP69K protection) is pre-mounted on the indicator.
Please use extreme caution when performing any work on the equipment that
affects this cable gland.
You must use a torque wrench. The torque for this cable gland is 5 Nm.