Operating Instructions Combics CAIXS2 99
Data interface
Data Interfaces
The indicator is equipped with the following data interfaces:
– COM1: –Standard data interface (RS-232 (A21), 485 (A22), 422 (A23).
The interface can be configured in the Setup menu for different input and output
functions (e.g. printer, 2nd weighing platform, PC).
3
Warning when using third-party RS-232 connecting cables: the pin assignments
may not be compatible with Sartorius equipment.
Specifications
Serial interface: Interface operation: Full duplex
Level: COM1: RS-232 or RS-422
Connection to device: Weighing platforms
Connection via screw terminals in the housing,
cable routed into the housing via a cable gland.
Transmission rate: 150, 300, 600, 1200, 2400, 4800, 9600, 19200
baud (depending on the operating mode)
Number of data bits: 7, 8 bits
Parity: Space, odd, even, none (depending on the
operating mode)
Number of stop bits: 1 or 2
Handshake mode: Software (XON/XOFF), hardware (1 character after
CTS)
Protocols: SBI, XBPI-232, XBPI-485, SMA
Network address
4)
: 0, 1, 2, ..., 31
SBI: Manual data output: Without stability, after stability, configurable
printout
SBI: Auto data output: Without stability, at stability, at user-defined
intervals
SBI: Output format: 16 or 22 characters
Printout of application data: Configurable printout