34 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Clearing the Preload
Setup information Clearing the preload when weighing in legal metrology is only possible using
the “Zero at Power On" menu item.
The “Clear Preload" function (menu item 1.9.9) must be allocated to the
J key.
3
Once the preload has been deleted, the J key must be reallocated back to its
original function in the Setup menu, e.g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
t Remove the preload weight from the weighing platform.
J t Start the “Clear Preload".
y After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.