34 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Clearing the Preload
Setup information – Clearing the preload when weighing in legal metrology is only possible using
the “Zero at Power On" menu item.
– The “Clear Preload" function (menu item 1.9.9) must be allocated to the
J key.
3
Once the preload has been deleted, the J key must be reallocated back to its
original function in the Setup menu, e.g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
t Remove the preload weight from the weighing platform.
J t Start the “Clear Preload".
y After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.