90 Operating Instructions Combics CAIXS2
Operation
Combining Application Programs
The following table shows how the applications described can be combined. The
basic weighing function is available at all times; it does not need to be combined
with a computational function.
Select programs one after the other: Toggle using the D key.
Application 1 Application 2 Application 3
(Basic Function) (Monitoring Function) (Cumulative-value Function)
Counting – Totalizing
Counting Checkweighing Totalizing
Counting Checkweighing –
Counting Classification –
Neutral measurement – Totalizing
Neutral measurement Checkweighing Totalizing
Neutral measurement Checkweighing –
Neutral measurement Classification –
Animal weighing – Totalizing
Animal weighing Checkweighing Totalizing
Animal weighing Checkweighing –
Animal weighing Classification –
Weighing in percent – Totalizing
Weighing in percent Checkweighing Totalizing
Weighing in percent Checkweighing –
Weighing in percent Classification –
– – Net-total formulation
– Checkweighing Totalizing
Example: “Portioning" (counting z, checkweighing O with totalizing s)
Configuration:
Application 1: Counting (COUNT.)
Application 2: Checkweighing (CHECK.)
Application 3: Totalizing (TOTALIZ): Saved value: Net + Calculated (3.23.3)
Autosave: On (3.16.2)
Source of data: Application 2 (3.22.2)
Setup: Printout: PRT PROT 7.8. Printer 1: “Total: Print when FN pressed,” then
select the menu line items of your choice.
t Place empty container on the scale.
) t Tare scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.