38 Operating Instructions Combics CAIXS2
Operation
Adjustment/Configuration Counter for Standard cales
Purpose Automatically record changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the device.
t To display both counters, press and hold the ( key for longer than
2 seconds.
y The “Configuration counter" is first of all shown in the weight display for
3 seconds (identified by a P). The “Adjustment counter" is then displayed for
another 3 seconds (identified by a C). After 6 seconds, the information display
turns off automatically.
Adjustment counter features: Counter limited to 9999
Counter at “C 0000" for hardware commissioning
Counter cannot be reset
Counter is updated automatically when:
linearization, calibration/adjustment is successful
user calibration, adjustment or linearization weight is changed (menu 1.18.)
when the following parameters are changed:
function of the q key (menu item 1.9.)
zero setting range (menu item 1.11)
tare/zero at power on (menu item 1.12)
the above parameters are reset to factory settings (menu item 9.1.1).
Configuration counter features: Counter limited to 9999
Counter at “P 0000" for hardware commissioning
Counter cannot be reset
Counter is updated automatically when:
the following parameters are changed:
installation location (menu item 1.1.)
application filter (menu item 1.2.)
stability range (menu item 1.3.)
taring (menu item 1.5)
auto zero (menu item 1.6.)
weight unit 1 (menu item 1.7.)
weight unit 2 (menu item 3.1.)
weight unit 3 (menu item 3.3.)
the above parameters are reset to factory settings (menu item 9.1.1)
switching the k key to or from a 10-fold higher resolution
turning the application automatic taring on/off
(menu item 3.7.)
the application parameters are reset to factory settings (menu item 9.1.1).