Operating Instructions Combics CAIXS2 103
Data interfaces
Example: Output weight value of +1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 5 5 . 7 * g * * CR LF
Position 1: Plus +, or minus - or space
Position 2: Space
Positions 3-10: Weight value with decimal point. leading zeros are output as
spaces.
Position 11: Space
Positions 12-14: Characters for unit of measure, space, or ! sign as a symbol
Position 15: Carriage return
Position 16: Line feed
Data Output Format with 22 Characters
Normal Operation
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
I I I I I I + D D D D D D D D D * U U U CR LF
I I I I I I - D D D D D D D D D * U U U CR LF
* * * * * * * * * * * * * * * * * * * * CR LF
I: ID code character, right-justified with spaces
+-: Plus or minus sign
*: Space
D: Digit or letter (max. 7 characters plus decimal point)
U: Unit symbol (1 to 3 letters followed by 2-0 spaces)
CR: Carriage return
LF: Line feed
Special Outputs
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * * * * - - * * * * * * CR LF
S t a t * * * * * * * * H * * * * * * * CR LF
S t a t * * * * * * * * H H * * * * * * CR LF
S t a t * * * * * * * * L * * * * * * * CR LF
S t a t * * * * * * * * L L * * * * * * CR LF
S t a t * * * * * * * * C * * * * * * * CR LF
*: Space
ā ā: Final readout
H: Overload
HH: Overload in checkweighing
L: Underweight
LL: Underweight in checkweighing
C: Adjustment