EasyManua.ls Logo

Sartorius Combics 2 CAIXS2 - Printer Interface Configuration (Printer)

Sartorius Combics 2 CAIXS2
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Combics CAIXS2 103
Data interfaces
Example: Output weight value of +1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 5 5 . 7 * g * * CR LF
Position 1: Plus +, or minus - or space
Position 2: Space
Positions 3-10: Weight value with decimal point. leading zeros are output as
spaces.
Position 11: Space
Positions 12-14: Characters for unit of measure, space, or ! sign as a symbol
Position 15: Carriage return
Position 16: Line feed
Data Output Format with 22 Characters
Normal Operation
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
I I I I I I + D D D D D D D D D * U U U CR LF
I I I I I I - D D D D D D D D D * U U U CR LF
* * * * * * * * * * * * * * * * * * * * CR LF
I: ID code character, right-justified with spaces
+-: Plus or minus sign
*: Space
D: Digit or letter (max. 7 characters plus decimal point)
U: Unit symbol (1 to 3 letters followed by 2-0 spaces)
CR: Carriage return
LF: Line feed
Special Outputs
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * * * * - - * * * * * * CR LF
S t a t * * * * * * * * H * * * * * * * CR LF
S t a t * * * * * * * * H H * * * * * * CR LF
S t a t * * * * * * * * L * * * * * * * CR LF
S t a t * * * * * * * * L L * * * * * * CR LF
S t a t * * * * * * * * C * * * * * * * CR LF
*: Space
– –: Final readout
H: Overload
HH: Overload in checkweighing
L: Underweight
LL: Underweight in checkweighing
C: Adjustment

Table of Contents