104 Operating Instructions Combics CAIXS2
Data interfaces
Error Message
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * U r r * * # # * * * * CR LF
S t a t * * * * * U r r * # # # * * * * CR LF
*: Space
#: Error code number (2 or 3 digits)
G# Gross value
N Net value
T Application tare memory 1
T2 Application tare memory 2
Diff Difference from adjustment value
Targ. Exact adjustment weight value
Nom. Exact adjustment weight
for SBI printout
nRef Reference sample quantity
pRef Percentage of reference
wRef Reference piece weight
Qnt Result from “Counting" (piece count) and “Neutral Measurement"
applications
mDef Target value for animal weighing
x-Net Animal weighing results
Setp Target value for checkweighing
Diff.W Absolute difference (e.g., in kg) in Checkweighing
Lim Deviation in % in Checkweighing
Max Upper tolerance for checkw.
Min Min. tolerance for checkw.Stat Status
Classx Classification
Limx Class limit
D Percentage (as loss)
Prc Percentage (as residue)
Wxx% Reference percentage weight
Cmpxxx Component xxx
Cont.T Contents of the tare memory in Net-total Formulation
S-Comp Total of initial weighings for Net-total Formulation
PT2 Preset tare
n Transaction counter
*G Sum of gross weights in Totalizing
*N Sum of net weights in Totalizing
Ser.no Serial number of the platform or indicator
Example: Output of the weight value +1255.7 g
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
G # * * * * + * * * 1 2 5 5 . 7 * g * * CR LF
Positions 1-6: ID code, right-justified with spaces
Position 7: Plus +, or minus - or space
Position 8: Space
Positions 9-16: Weight value with decimal point. leading zeros are output as
spaces (a comma can also be set instead of a decimal point,
menu item 7.17)
Position 17: Space
Positions 18-20: Characters for unit of measure, space, or ! sign as a symbol
Position 21: Carriage return
Position 22: Line feed