24 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Analog/Digital Converter (ADC)
Purpose Adjust the parameters of the analog/digital converter to the connected load cell or
weighing platform. After ADC configuration, the ADC in connection with the load
sensor is defined as a scale.
3
Once the ADC configuration has been locked, the indicator can no longer be used
to influence weighing results. The scope of functions available in the weighing
instrument is defined by the A/D converter. Weighing functions that can be
activated include reading weight values, taring, adjustment, reading the tare
value and saving/deleting the tare entry.
Setup information ADC configuration is only possible when the menu access switch is open. Close
the menu access switch after ADC configuration, as otherwise there will not be
any display of the conditions “overload" ( H ) and “underload" ( L ).
Before ADC configuration, you must first set whether or not the weighing
platform will be used as a standard or verifiable weighing platform under menu
item 9.1.
When the Service mode is activated, the ADC configuration takes place in the
Setup menu under WP-1 for the first weighing platform and under COm1 /
WP-2 for the second weighing platform.
3
If you return to the highest level of the Setup menu without saving the
configuration parameters beforehand (menu item 11.10) any settings that have
been made will be deleted.
The settings are made in the corresponding Setup menu under menu item 11.
Enter the maximum capacities in a suitable weight unit, without any decimal
places (decimal places will be truncated by the rounding function).
Entries made in the ADC configuration will not be affected by a menu reset
(returning the setup parameters to their factory settings).
Factory settings/Reset menu 9.1
wt.para
STANDRD Standard configuration 9.1.3
RANGE Ranges 11.3
SINGLE Single-range scale 11.3.1
MULT.INT Multi-interval scale 11.3.2
MULT.RNG Multiple-range scale 11.3.3
SINGLE Single-range scale 11.4
d Scale interval d 11.4.1
MAX Max. load 11.4.4
MULT.INT Multi-interval scale 11.5
d Scale interval d 11.5.1
RANGE 1 Range 1 11.5.4
RANGE 2 Range 2 11.5.5
RANGE 3 Range 3 11.5.6
max Max. load 11.5.7
MULT.RNG Multiple-range scale 11.6
d Scale interval d 11.6.1
RANGE 1 Range 1 11.6.4
RANGE 2 Range 2 11.6.5
RANGE 3 Range 3 11.6.6
MAX Max. load 11.6.7
WT.UNIT Available weight units 11.7
FREE User-defined /o 11.7.1
g Grams /g 11.7.2
kg Kilograms/kg 11.7.4
t Tons/t 11.7.21
lb Pound:ounces/lb o 11.7.22