Operating Instructions Combics CAIXS2 71
Operation
CTRL.SET Activate SET control output 4.3
OUTPUT “SET" output 4.3.1*
OP.READY Ready to operate 4.3.2
OUTP.ACT Port lines 4.4
OFF Off 4.4.1
ALWAYS Always on. 4.4.2
STABIL On at stability 4.4.3
CHECK.RG On within checkweighing range 4.4.4*
STAB.CHK On at stability within checkweighing
range 4.4.5
INPUT Parameter input 4.5
TAR.MN.MX Min, Max, target value 4.5.1*
TARG.PER Only target value with percent limits 4.5.2
TAR.A.PER Target value with asymmetrical percent
limits 4.5.3
TAR.TOL Target value with relative tolerances 4.5.4
AUT.PRNT Automatic printing 4.6
OFF Off 4.6.1*
ON On 4.6.2
OK Only values within tolerance 4.6.3
NOT OK Only values outside tolerance 4.6.4
APP.ZERO Checkweighing toward zero 4.7
OFF Off 4.7.1*
ON On 4.7.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Display The result of a measurement is shown either as a weight value or in relation to the
target.
Weight display: The measured value line always shows the weight value, even if it
lies outside the tolerance range. The bar graph is displayed with symbols indicating
lower limit, target and upper limit. Weights are shown logarithmically from 0 up to
the lower tolerance limit, and linearly beyond that point.
Relation to target value: As “Weight display" above, with the exception that:
LL appears in the main display if the weight value is less than the lower limit
HH is shown on the main display if the weight value is higher than the upper
tolerance limit.
Digital I|O Interface The Checkweighing application supports the digital input/output interface.
The four outputs are activated as follows:
Less than > red LED lights up
Equal to > green LED lights up
Greater than > yellow LED lights up
– Set.
The outputs can also be galvanically isolated using option A5.
Acoustic signal: An acoustic signal can be activated in addition to the green LED.
Setting: menu item 8.2.3.