Operation
Entering the Reference Piece Weight
The reference piece weight (i.e. the weight of one piece) can be entered using the
keypad and saved with the O key.
h
The entered value remains active until deleted by pressing the c key or
overwritten by a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC. / APPLIC.1 / COUNT. menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1-digit 1 scale interval* 3.6.1
2 digit 2 scale intervals 3.6.2
5 digit 5 scale intervals 3.6.3
10 dig. 10 scale intervals 3.6.4
20 dig. 20 scale intervals 3.6.5
50 dig. 50 scale intervals 3.6.6
100 dig. 100 scale intervals 3.6.7
200 dig. 200 scale intervals 3.6.8
500 dig. 500 scale intervals 3.6.9
1000 d 1000 scale intervals 3.6.10
RESOLUT Resolution for calculation of reference value 3.9
Disp.acc. Display accuracy* 3.9.1
10 FOLD Display accuracy + 1 decimal place 3.9.2
100 FOLD Display accuracy + 2 decimal places 3.9.3
SAVE WT.Parameter for saving weight values 3.11
STABIL With stability* 3.11.1
ACC.STAB With increased stability 3.11.2
REF.UPDT Reference sample updating 3.12
OFF Off 3.12.1
AUTOMAT Automatic* 3.12.3
REF.WP Reference weighing instrument 3.13
NO WP No weighing platform selected 3.13.1*
WP 1 Weighing platform WP1 3.13. 2
WP 2 Weighing platform WP2 3.13. 3
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e., the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
error code Inf 29 appears
the weighing platform is not initialized
the preset reference sample quantity is saved.
Setting: APPLIC./APPLIC.1/COUNT./MIN.INIT menu item 3.6.
The minimum load can be set in 10 steps, from 1 to 1000 digits (see Available
parameters). The “digits" here refer to the scale intervals in the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and
1000 digits are required, you must place at least 1000 g (= 1000 intervals =
1000 digits) on the weighing platform for initialization.