Operation
Manual start with user-defined number of subweighing operations: Place the
sample on the platform and enter the number of weighing operations using the
keypad. Press the r key to save the number entered and begin weighing.
Automatic start with preset number of subweighing operations: Measurement
begins when you place the first sample on the platform, provided the start
conditions are met.
Preparation t Open the APPLIC./APPLIC.1/ANIM.WGmenu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1-digit 1 scale interval* 3.6.1
2 digit 2 scale intervals 3.6.2
5 digit 5 scale intervals 3.6.3
10 dig. 10 scale intervals 3.6.4
20 dig. 20 scale intervals 3.6.5
50 dig. 50 scale intervals 3.6.6
100 dig. 100 scale intervals 3.6.7
200 dig. 200 scale intervals 3.6.8
500 dig. 500 scale intervals 3.6.9
1000 d 1000 scale intervals 3.6.10
START Start averaging 3.18
manual Manual* 3.18.1
AUTOMAT Automatic 3.18.2
ACTIVTY Animal activity 3.19
0.1 perc. 0.1% of animal/object 3.19.1
0.2 perc. 0.2% of animal/object* 3.19.2
0.5 perc. 0.5% of animal/object 3.19.3
1 PERC. 1% of animal/object 3.19.4
2 PERC. 2% of animal/object 3.19.5
5 PERC. 5% of animal/object 3.19.6
10 PERC. 10% of animal/object 3.19.7
20 PERC. 20% of animal/object 3.19.8
50 PERC. 50% of animal/object 3.19.9
100 PERC. 100% of animal/object 3.19.10
PRINT Autom. printout of results 3.20
manual Off* 3.20.1
automat On 3.20.2
DIS.UNLD Static display of result after load removed 3.21
CLEARED Display is fixed until 3.21.1
unload threshold reached*
PRESENT Fixed display until c is pressed 3.21.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e., the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
error code Inf 29 appears
the weighing platform is not initialized
the preset reference sample quantity is saved.
Setting: Applic./Applic.1/ANIM.WG/Min.init menu item 3.6.