Operating Instructions Combics CAIXS2 65
Operation
Entering the Reference Percentage Value
The reference weight for 100% is entered using the keypad and the O key is
pressed to initialize the application.
h
The entered value remains active until deleted by pressing the c key or
overwritten by a new value. It remains saved after the scale is switched off.
Preparation t Open the APPLIC./APPLIC.1/PERCENT menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1-digit 1 scale interval* 3.6.1
2 digit 2 scale intervals 3.6.2
5 digit 5 scale intervals 3.6.3
10 dig. 10 scale intervals 3.6.4
20 dig. 20 scale intervals 3.6.5
50 dig. 50 scale intervals 3.6.6
100 dig. 100 scale intervals 3.6.7
200 dig. 200 scale intervals 3.6.8
500 dig. 500 scale intervals 3.6.9
1000 d 1000 scale intervals 3.6.10
RESOLUT Resolution for calculation of reference value 3.9
Disp.acc. Display accuracy 3.9.1*
10 FOLD Display accuracy + 1 decimal place 3.9.2
100 FOLD Display accuracy + 2 decimal places 3.9.3
Dec.plcs Decimal places in displayed result 3.10
without none 3.10.1*
1 Dec.Pl. 1 decimal place 3.10.2
2 Dec.Pl. 2 decimal places 3.10.3
3 Dec.Pl. 3 decimal places 3.10.4
SAVE WT.Parameter for saving weight values 3.11
STABL. With stability 3.11.1*
ACC.STAB With increased stability 3.11.2
REF.WP Reference weighing instrument 3.13
NO WP No weighing platform selected 3.13.1*
WP 1 Weighing platform WP1 3.13.2
WP 2 Weighing platform WP2 3.13.3
CALC.DIS Calculated values display 3.15
RESIDUE Residue 3.15.1*
LOSS Loss 3.15.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Minimum load for initialization You can set the minimum load here, i.e., the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
error code Inf 29 appears
the weighing platform is not initialized
the preset reference sample quantity is saved.