Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 56
Printer LPT1: (if existing)
assigned: The interface is not enabled for a printer in the menu interfaces.
off: No printing
Test & print: The test results will be printed.
Self-test: The self-test will be printed.
Test & Selftest: Both, the test results and the result of the self-test will be printed.
Print all: The results printed after the test program has been completed. It is possible to
enter the desired number of print-outs.
Print all go: The results printed after the test program has been completed and the test is
OK (go). It is possible to enter the desired number of print-outs.
Print all no go: The results printed after the test program has been completed and the test is
not OK (no go). It is possible to enter the desired number of print-outs.
Header:
on: The results are printed together with a header.
off: The results are printed without a header.
Definitions:
on: The results are printed with their complete definition (40 characters). The
printout will therefore have DIN A4 format.
off: The results are printed without definition. The printout will be considerably
smaller.
Protocol number
Here you can determine, whether the printout of the protocol is to contain a serial number that has been
defined before. The protocol number will automatically be increased by 1.
on: The printout will contain a protocol number.
off: The printout will contain no protocol number.
Protocol number
(Ab Version 4.308 im Menü General Data)
The protocol number to appear on the printout is set here.
Increm. Counter
(Ab Version 4.308 im Menü Allgemeine Daten)
Do not count: On the printout, the preset protocol number will remain the same.
Count go: “go” test processes will increase the counter of the protocol number by 1.
Count no go: “no go” test processes will increase the counter of the protocol number by 1.
Count go / no go: All test processes will increase the counter of the protocol number by 1.
Print protocol (only in connection with the optional clock)
With date: The protocol will contain a date.
With time/date: The protocol will contain time and date.
Without time/date: The protocol will contain neither time nor date.
Free columns
On the paper, the printout on the external printer can be moved to the right by entering blanks. You need
to enter the number of blanks.
Free lines
On the paper, the printout on the external printer can be moved to the bottom by entering free lines. You
need to enter the number of free lines.