Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 *
http://www.schleich.com * info@schleich.com 65
Out of memory! Test program can't be stored!
The present test program cannot be stored, because the storage of the internal database is too small. More
storage capacity can be created by deleting some test programs.
Error in data record. Can't be loaded!
This test program in the internal storage is defective and cannot be loaded. Please contact the manufacturer
of the testing device.
Error at voltage control (fuse broken?)
For the function test or for the leakage current test, some testing devices are equipped with a regulating
transformer, which can be adjusted automatically or manually.
If adjusted manually, a window will appear during the test, indicating the set value and the actual value
of the voltage. Now, the operator can adjust the voltage manually. If the voltage is correct, the test will be
continued automatically. If the voltage is not adjusted within 30s, you will see the above error message. If
the actual value of the voltage is very small, it is possible that the fuse of the function test has opened.
If adjusted automatically, there will also be a window indicating the set value and the actual value of the
voltage. If this voltage remains the same for 5s, this window appears. If the actual value of the voltage is
very small, it is possible that the fuse of the function test has opened.
Voltage is set automatically
The voltage of the function test is adjusted automatically. This error message appears in case the voltage
control has produced an error (see above).
Please, re-adjust the voltage!
For the function test, the testing device waits for the manual adjustment of the test voltage. If an error
occurs, the above message appears.
Below minimum current at leakage current test. Cancelled!
It is possible to enter a minimum current for the leakage-current test. If this minimum current is fallen
below during the test, the above error message appears. If the test object has no leakage current, this
setting must be 0. If the test object usually has a leakage current, there could be a contacting problem in
this case.
First enter name of user
Before the test can be started, it is necessary to read in a user name with the bar-code scanner. This has
been set in the menu “General data” under “Bar code input”.
First enter serial number
Before the test can be started, it is necessary to read in a serial number with the bar-code scanner. This has
been set in the menu “General data” under “Bar code input”.
First enter name of test program
Before the test can be started, it is necessary to read in a test program with the bar-code scanner. This has
been set in the menu “General data” under “Bar code input”.
The label program is being sent to the printer. Please wait.
The software is currently transmitting a label program to the label printer. Please wait.
The label program is being transmitted. Please wait.
The software NetCom is currently transmitting a label program to the tester. Please wait.
Label print-out. Please wait.
Currently data is being transmitted to the label printer. Please wait.