SciCan STATCLAVE G4 - Water Supply; Safety Information

SciCan STATCLAVE G4
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Water supply
High quality distilled water is recommended for use in the STATCLAVE G4. Deionized, demineralized or
specially filtered water can also be used. Never use tap water. The STATCLAVE G4’s water conductivity sensor
automatically reads the water quality and will not allow the unit to run a cycle unless the water quality is
6.4ppm / 10 µS (conductivity at 20˚C/ 68˚F).
Safety information
THE FOLLOWING TERMS APPEAR IN THIS SERVICE MANUAL
CAUTION!
A potential hazard to the operator or to patients.
IMPORTANT!
A situation that may affect the functioning of the unit or lead to a mechanical failure.
TIP
Additional information that may be helpful.
Pay close attention to the following symbols that appear on the unit: