85
Troubleshooting loading errors and operator complaints
Machine will not start, and
touchscreen shows: [H2OX]
Clean water reservoir issue
either water level is too low or
water used is of inadequate
quality.
Press on the icon to confirm
whether it is a water level
problem or a water quality
problem.
If water level is too low: Refill
the reservoir. Refer to the steps
described in section 3.4 Filling
the Reservoir.
inadequate: You have likely
used water that is not steam-
process distilled or is improperly
distilled.
Empty the reservoir and refill
with steam-process water
containing less than 5 ppm total
dissolved solids (having
conductivity of less than 10 μS /
cm).
If you have a water conductivity
meter, check the quality of the
water before refilling the
reservoir.
To empty the reservoir, see
section XX Shipping the unit.