EasyManua.ls Logo

SciCan STATCLAVE G4 - Safe operation

SciCan STATCLAVE G4
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Safe operation
The following apply to both operators and service technicians:
Exercise caution and seek assistance when lifting or carrying the unit.
Before performing routine maintenance or servicing the unit, turn the unit OFF and unplug the power
cord from the power source.
The operator should never remove the cover of the unit or insert objects through holes or openings
in the cabinetry. Doing so may damage the unit and/or pose a hazard to the operator.
If the unit is used in a manner other than that specified, the protection provided by the equipment
may be impaired.
Safe servicing
SciCan shall not be liable for incidental, special or consequential damages caused by any maintenance
or services performed on the STATCLAVE G4 by a third party or for the use of equipment or parts
manufactured by a third party, including lost profits, any commercial loss, economic loss, or loss
arising from personal injury.
All local, regional, state, and national regulations regarding the servicing of this class of device and
safety requirements must be observed.
When the panels are removed
Hazardous voltages are accessible. Disconnect the power cord before removing any panels.
Sharp metal edges are exposed. Be careful and wear long sleeves and gloves.
Electrical
If the panels are removed, a dielectric strength test (Hi-Pot) AND a protective bonding impedance
test (ground continuity) must be performed on the STATCLAVE G4 when the work is completed and
after the panel has been reinstalled.
The STATCLAVE G4 contains electronic circuitry that is static sensitive. Always wear a static strap
when working with or near printed wiring boards. In addition, use static footstraps, grounding mats
and grounded work surfaces when servicing microprocessor devices. Transport boards and devices
in static protected bags.
Lifting the unit
The STATCLAVE G4 is heavy. Exercise caution and seek assistance when lifting or carrying the unit.
Water quality
Use only steam-process distilled water in the STATCLAVE G4.
Ensure that there is distilled water in the STATCLAVE G4 before activating the pump.