80
Step-by-step troubleshooting by symptom
issue
Check that the unit is plugged into a
properly grounded outlet and that the
power cord is firmly seated at the rear
of the machine.
Try another outlet. Power unit OFF for
10 seconds and then power ON again.
Check the condition of the line circuit
breaker or fuse.
machine.
Spill over from refilling
External water tank feed
issue
Check that water was not spilled when
refilling the reservoir.
Check that the tube coming from the
external tank (if fitted) is completely
pushed onto the connector.
Run a Vacuum Test. If water drips from
the underside of the unit during the
test, call your SciCan dealer.
STERILE, Cycle aborted — NOT
STERILE and CYCLE FAULT
messages.
pressed while the unit was in
operation.
A power outrage or power
fluctuation occurred while
the unit was in operation.
Wait a few minutes and attempt
another cycle before proceeding to the
next solution.
NOTE: STATCLAVE G4 units connected
to the Internet and registered with
SciCan will automatically send Cycle
Fault messages to SciCan’s
international service center.
Excessive steam issuing from
the front of the machine.
Open and close the door then attempt
another cycle. Check the door seal for
misalignment or damage. Replace the
door seal if required.
If the leak persists, turn the unit OFF,
remove the load and contact your
SciCan dealer.