62
Annual service recommendations
Clean reservoirs and reservoir filters
Check integrity of incoming and outgoing services (power, water supply, drain)
Check general condition of machine
Check bacteriological filter (replace if required)
Inspect and clean right cover air filter
Check door seal (replace if required)
Inspect chamber filters (replace if required)
Software upgrade (if required)
Check the pressure relief valve and power failure pressure/vacuum relief
Calibrate the unit if necessary (See Calibration section for details on when calibration is
needed) or in accordance to local guidelines