32
Test cycles
Running a vacuum test
The vacuum test checks the sterilizers plumbing system for leaks. Run this test with the rack and empty
trays in the chamber.
IMPORTANT: Vacuum tests must be conducted when the unit is cool. Running a vacuum test on a hot
chamber may cause the test to fail.
Running a vacuum test can take a minimum of 45 minutes. When the test is complete, the screen will
display a CYCLE COMPLETE message. If the test has failed, see Troubleshooting.