Using the service menu
To access the service menu, select the SETTINGS icon from the home screen, press on the image of the
technician and enter the service code 7919 on the keypad.
Menu item
What to do with it
PRODUCTION TOOLS
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.
DIAGNOSTIC TOOLS
This menu contains a number of tools technicians can use to more accurately diagnose problems.
Use to run a series of cycles in
succession. Can be set to repeat
1 to 5 cycles in a row or set it to
ON to run the same cycle over
again until you stop it.
To display the debug screen
while running a cycle, set Debug
Screen to ON.
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.